Research on Measurement of Human Ocular Off-Axis Aberrations Based on Hartmann-Shack Sensor

XU An-cheng, LU Xing-Zhong, DING Li-hua   

  1. (Changzhou Instituted of Technology, Changzhou 213002, China)
  • Published:2013-01-28 Online:2013-01-07

摘要:

人眼是一个不完善的成像系统,存在着各种像差,包括轴上像差及轴外像差, 这些像差都会影响视网膜的成像质量。而现阶段大多数人眼像差的检测方法主要针对轴上视场像差的测量,比如离焦、像散以及部分高阶像差的测量。因此提出了基于Hartmann-Shack波前传感器的人眼离轴像差的测量方法,建立测量系统并进行实验研究,当离轴角度从0°偏离到左右各5°、10°时,均方根(RMS)值分别从1.12变化到1.36(1.52),1.41(1.96),结果表明随着离轴角度的增加人眼外围视野的成像质量显著下降,原因一方面是人眼离轴像差的增大,另一方面是由于外围视野的感光细胞密度降低。

关键词: 波前, 离轴像差, Hartmann-Shack传感器

Abstract:

The human eye is an imperfect refractive system which has kinds of aberrations, including on-axis aberrations and off-axis aberrations, all of which have influence on the imaging quality of retina. But at present most of the ocular aberration detection methods focus on on-axis measurement, such as defocus, astigmatism, and part of high orders aberrations. The ocular off-axis aberrations detection method based on the Hartmann-Shack is proposed, the measuring system is set up and the experimental test is carried out. When the off-axis angle changes from 0° around to 5°, 10°, root mean square(RMS) values change from 1.12 to 1.36 (1.52), and 1.41 (1.96), and the results show that with the increase of the off-axis angle, the image quality of human peripheral vision decreased significantly. The reason is the increase of the off-axis aberrations, the other reason is due to the decrease of the peripheral vision photoreceptor cells density.

Key words: wavefront, off-axis aberrations, Hartmann-Shack sensor