Journal of Atmospheric and Environmental Optics ›› 2009, Vol. 4 ›› Issue (5): 393-400.

• 论文 • Previous Articles    

Optical Signal Detecting and Processing For Water Quality Measuring System by Ultraviolet Spectrum Absorption

CHENG Yin, LIU Jian-guo, LU Yi-huai, WANG Ya-ping, FANG Wu, CHEN Jun   

  1. (Key Lab of Environmental Optics & Technology, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Heifei 230031,China)
  • Received:2009-05-05 Revised:2009-08-17 Online:2009-09-28 Published:2009-09-28

Abstract:

In water quality measuring methods using ultraviolet spectrum absorption, optical signal detecting techniques are very important. Key techniques are the time-sharing sampling system containing background light sampling, referring and measuring with one detector in double light path, and eliminating influence that comes from the drift of light source and background light. The double wavelength band detecting model can overcome the influence of turbidity in water. The techniques make apparatus more precise and reliable. The standard deviation is less than 1% and the maximum relative error is less than 2% for measurement of standard solutions. These results indicate that the precision and accuracy of the apparatus are good.

Key words: water pollution, absorbance, ultraviolet absorbance, signal detecting and processing

CLC Number: