Journal of Atmospheric and Environmental Optics ›› 2013, Vol. 8 ›› Issue (3): 226-231.

• 论文 • Previous Articles     Next Articles

Micro-Thermal Meter Based on Two-channel Constant Temperature Wind Tunnel System

YAO Yuan-cheng1, 2, WENG Ning-quan1,3*, SUN Gang1, LIU Qing1, ZHANG Cai-yun1, 2, XIAO Li-ming1   

  1. (1 Key Laboratory of Atmospheric Composition and Optical Radiation, Anhui Institute of Optics and Fine Mechanics, Chinese Academy of Sciences, Hefei 230031, China; 
    2 University of Chinese Academy of Sciences, Beijing 100039, China; 
    3 School of Environmental Science and Optoelectronic Technology, University of Science and Technology of China, Hefei 230031, China)
  • Received:2013-03-06 Revised:2013-03-19 Online:2013-05-28 Published:2013-05-17

Abstract:

The design and performance of two-channel constant temperature wind tunnel detection system is introduced. The tunnels of system can make temperature and wind speed stable, and the differential temperature can be adjusted. The performance of the system is tested through experiments. The standard deviation of air temperature fluctuation of the fixed points in wind tunnel (0.025℃-the maximum) is smaller than which ( 0.059℃)in room. The transverse field of temperature in wind tunnel is not homogenous, which in cold and heating tunnel the standard deviations are 0.044℃ and 0.059℃. The system is used as a platform mainly for testing HTP-2 micro-thermal meter. It is proved that heat radiation has bad influence on measurement of micro-thermal meter by the system, and the radiative heating of the wire filament of micro-thermal meter is 0.365℃.

Key words: atmospheric optics, constant temperature wind tunnel, micro-thermal meter, heat radiation

CLC Number: