Journal of Atmospheric and Environmental Optics

Previous Articles     Next Articles

On-Orbit Radiometric Calibration of Spaceborne Differential Optical Absorption Spectrometer Based on the Sun

WANG Shimei, SI Fuqi, ZHAO Minjie, ZHOU Haijin, JIANG Yu   

  1. Key  Laboratory of  Environmental  Optics  and  Technology,
    Anhui  Institute  of  Optics  and  Fine  Mechanics,  Chinese  Academy  of  Sciences,\quad Hefei 230031, China
  • Online:2019-01-28 Published:2019-01-29

Abstract: On-orbit radiometric calibration of spaceborne differential optical absorption spectrometer can be conducted based on the bi-directional scattering distribution function (BSDF) and the sun. The EMI BSDF is the ratio of EMI radiance and irradiance calibration function. Radiometric calibration of observation path is completed on the ground. Radiometric calibration of calibration path can be realized by the sun. Then EMI BSDF is obtained, and the on-orbit radiometric calibration is conducted. The results between the EMI and TROPOMI show that the accuracy of the radiance calibration is 5%~6.5%.

Key words: spaceborne imaging spectrometer, bi-directional scattering distribution function, on-orbit radiometric calibration

CLC Number: